Backed by proprietary Swiss sensor technology and materials, the LabSen 831 HF pH Electrode Probe is designed for pH measurements for solutions containing Hydrofluoric acid (≥3pH), or other strong acid solutions.
Special Glass Membrane – Unique LabSen HF glass membrane, which has a faster response rate, better repeatability in strong acid samples, and resistant to hydrofluoride corrosion.
Silver Ion Trap Reference System – The LabSen 831 pH electrode adopts a Silver Ion Trap reference system. On top of the Long-Life reference system, a silver ion trap is added to prevent the junction being blocked by silver sulfide.
The electrode uses a BNC connector, compatible with any pH meters that use BNC connectors, such as Apera Instruments MP511, SX811, PH700, PH800, PC800, PH820 and PC820. Temperature sensor is not built-in. Users need to have a separate temperature probe (ex: Apera Instruments MP500) to enable ATC or manually input temperature to realize temperature compensation.
The electrode is backed up with a 6-month warranty by Apera Instruments.
Code |
AI3109 |
|
Material |
Lead-free glass |
|
Built-in Temperature Compensation Sensor |
No |
|
Junction |
Ceramic |
|
Reference |
Silver Ion Trap |
|
Electrolyte |
3M KCL |
|
Length |
120 mm |
|
Diameter |
12 mm |
|
Membrane Shape |
Half-Ball |
|
Range |
0-11 pH |
|
Temp. Range |
0 to 100˚C (32 to 212˚F) |
|
Connector |
BNC |
|
Cable |
3 ft |