C106H Gas Permeability Test System is based on the differential pressure method, and is professionally applicable to the determination of gas transmission rate, solubility coefficient, diffusion coefficient and permeability coefficient of plastic films, composite films, high barrier materials, sheets, and metal foils at different temperatures. The test system conforms to GB, ISO, ASTM and other international standards.
Accurate Data
High Efficiency
Intelligent control
Safe and reliable
Space saving
Powerful functions
The pre-conditioned specimen is mounted in the test cell as to form a sealed barrier between two chambers. The lower-pressure chamber (lower chamber) is firstly evacuated, followed by the evacuation of the entire cell. A flow of gas is thereafter introduced into the evacuated higher-pressure chamber (upper chamber) and a constant pressure (adjustable) difference is generated between the two chambers. The gas permeates through the specimen from higher pressure side into the lower pressure side. The gas permeability and other barrier properties of the specimen can be obtained by analyzing the pressure changes in the lower chamber.
ISO 15105-1、ISO 2556、GB/T 1038-2000、ASTM D1434、JIS K7126-1、YBB 00082003
Applications |
---|
|
|
Table 1: Test Parameters Note2
Parameters/Model | C106H | |
---|---|---|
Test range | cm3/m2·24h·0.1MPa | 0.01~50,000 |
Resolution | cm3/m2·24h·0.1MPa | 0.001 |
Test temperature | ℃ | 15~50,5~60(Optional) |
Temperature resolution | ℃ | 0.01 |
Temperature fluctuation | ℃ | ±0.15 |
Test humidity | %RH(Within standard test temperature range) | 0%,5~90%±2%(Customize) |
Vacuum Resolution | Pa | 0.01 |
Vacuum Accuracy | Pa | ±0.2%(1%-100% of Full Scale) |
Vacuum Degree of Test Chamber | Pa | ≤ 10 |
Additional Functions | DataShieldTM note3 | Optional |
GMP Computer System requirement | Optional | |
CFR21Part11 | Optional |
Table 2: Technical Specifications
Test Cell | 6 Cells |
---|---|
Specimen Size | 4.6” x 4.6”(11.7cm×11.7cm) |
Specimen Thickness | ≤120 Mil(3mm) |
Standard test area | 38.48cm2 |
Test Gas | O2、N2、CO2 etc.(Outside of supply scope) |
Test Pressure | 10 kPa~210 kPa(Can be set between) |
Pressure Fluctuation | 0.2 kPa |
Gas Pressure | 79.7 PSI / 550 kPa |
Port Size | Φ6 mm PU tubing |
Instrument Dimension | 23.6” H x 19.2” W x 25.9” D (60cm× 49cm× 66cm) |
Power Supply | 120VAC±10% 60Hz/220VAC±10% 50Hz(one of two) |
Net Weight | 220Lbs(100kg) |
Table 3: Product Configuration
Standard Configuration | Instrument mainframe, tablet, vacuum pump (UK), sampler, vacuum grease, Φ6 mm PU tubing |
---|---|
Optional Parts | Air compressor, CFR21Part11, GMP Computer System requirement, DataShieldTM note3 |
Note | The gas compressor port of the instrument is Φ6 mm PU tubing(pressure≥79.7 PSI/ 550kPa), customers need to prepare gas supply. |
Note 1: The described product functions are subject to the specification in "Technical Parameters" "Table 1: Test Parameters".
Note 2: The parameters in the table are measured in Labthink laboratory by professional operators according to the requirements and conditions stapulated in laboratory environmental standards.
Note 3: DataShieldTM provides safe and reliable data application support. Multiple Labthink instruments can share one single DataShieldTM system which can be configured as required.