The xenon light source is widely used in scientific research in the areas of hydrogen and oxygen photolysis, CO2 reduction, Synergy between light and heat, photothermal catalysis, photochemical catalysis, photochemical synthesis, photodegradable pollutants, water pollution treatment, biological lighting, optical inspection and solar simulation.


LMP400 lifting platform
Advantages of the CEL-PF300-T9 instrument
1. The CEL-PF300-T9 is an integrated power supply and light box.
2. The new power supply is integrated into the shooting module, significantly reducing the number of wires and considerably reducing the failure rate;
3. The trigger voltage of the new power supply is half of the original (±15KV), greatly reducing interference with the outside world;
4. The lamp uses the latest heat dissipation structure of the module to improve the stability of the light emission;
5. The PE300 lamp user can quickly replace it without any connection or disassembly;
6. Adopt the latest optical path structure to perform multiple lamp concentricity, heat dissipation, insulation, flange, etc.;
7. A new temperature control feedback system is used to isolate high voltage and improve accuracy;
8. The standard targeting lens is compatible with the entire range of M62 and M52 filters, as well as with any overlapping multi-layer filters;
9. The T9 light source can be equipped with the LMP400 electric lift platform to allow for a convenient lifting adjustment;
Technical Parameters
|
CEL-PF300-T9 |
CEL-PF300UV-T9 |
|
|
Output light power |
0~20 Sol (1 Sol = 1000 W/m²) |
0~20 Sol |
|
Spectral output (nm)
|
300nm~2500nm (free ozone) |
200nm~2500nm (ozone) |
|
Beam diameter |
Above 60 mm |
Above 60 mm
|
|
UV emission, <390nm (watts) |
2,6 W |
6,6 W |
|
Infrared output, >770nm (watts) |
28,8 W |
26,8 W |
|
Visible output, 390-770nm (lumens) |
5000Lu, 18,6W, 5600K |
4500Lu, 16,6W, 5050K |
|
Input power (Watts) |
300W (DC ignition) ±15KV, 14V DC operation) |
|
|
Working current (DC AMPIPERS) |
21A (10A~22A) |
|
|
TOTAL LIGHT OUTPUT POWER |
50W |
|
|
Lamp service life (hours) |
>1000H |
|
|
stability of the output light |
±1% |
|
|
Temperature control system |
Light source adopts multi-point temperature monitoring to ensure stable light source emission; The fan speed delay is automatically adjusted according to the system temperature to stabilize the emitted light intensity; |
|
|
Options |
CEL-LMP400 lifting platform Box for photochemical experiments CEL-LB70 CEL-NP2000-2 optical power gauge |
|
|
Optional filter with quartz coating (M62)
|
Available Optical Filters AREF (Total Reflection, 200-2500 Nm) - Standard included in the set of lamps VisREF (reflects visible light, 350-780nm) UVREF (reflect UV, 200-400nm) AM1.5G (300-1100nm) UVIRCut400 (pass 400-780nm) UVIRCut420 (pass track 420-780nm) CUT800 (passes above 800nm) Band Pass Filters QD254, QD275, QD280, QD295 QD300-800, QD313, QD320, QD325, QD330, QD334, QD350, QD355, QD360, QD365, QD370, QD375, QD380,QD400, QD405, QD420, QD435, QD450, QD475, QD500, QD520, QD550, QD578, QD600, QD630, QD650, QD670, QD700, QD730, QD765, QD850, QD940 |
|