The Vertex.One is an all-round potentiostat/galvanostat/ZRA with optional FRA/EIS. Its low cost and wide specifications result in a versatile single-channel instrument, and multiple instruments are an attractive alternative to a multi-channel solution. The cell is connected via our shielded cell cable, or standard 4mm banana leads allowing the user to supplement or extend the cell cable. The Vertex.One is capable of all standard electrochemical techniques and includes a complete suite of IviumSoft control and data processing software (for Windows based PCs).
The robust design, the wide specification range, the floating operation, and the availability of a wide range of electrochemical techniques makes the Vertex.One ideal for many applications, including:
The Vertex.One can optionally be expanded with integrated True Linear Scan module and/or FRA/EIS capability.
The Vertex.One is controlled via Ivium’s own IviumSoft. This versatile, yet intuitive software allows instrument control, data management and analysis, peripheral instrument control, etc. IviumSoft can also interface with and be controlled from LabVIEWTM, C++, Delphi, etc. A full suite of IviumSoft is included as standard with each instrument.
System performance |
|
Product Current compliance | ±100mA |
Product Maximum output voltage | ±10V |
Product Maximum compliance voltage | ±21V |
Product Electrode connections | 4; WE, CE, RE, S (and GND) with 4mm banana plugs |
Product Potentiostat bandwidth | >250kHz |
Product Stability settings | High Speed, Standard and High Stability |
Product Programmable response filter | 1MHz, 100kHz, 10kHz, 1kHz, 10Hz |
Product Signal acquisition | Dual channel 16bit ADC, 100,000 samples/s |
Potentiostat |
|
Product Applied potential range | ±10V, 0.333mV res. |
Product Applied potential accuracy | 0.2%, or 2mV |
Product Current ranges | ±100pA to ±100mA in 10 decades |
Product Measured current resolution | 0.015% of current range, minimum 0.01pA |
Product Measured current accuracy | 0.2% |
Galvanostat |
|
Product Applied current resolution | 0.033% of applied current range |
Product Applied current accuracy | 0.2% |
Product Measured potential resolution | 0.003% of potential range, minimum 400nV |
Product Measured potential accuracy | 0.2% or 2mV |
Impedance Analyser |
|
Product Frequency range | 10µHz to 250kHz |
Product Amplitude | 0.015mV to 1.0V, or 0.03% to 100% of current range |
Product DC offset | 16bit DC offset subtraction and 2 DC-decoupling filters |
Electrometer |
|
Product Input impedance | >1000Gohm //<10pF |
Product Input bias current | <10pA |
Product Bandwidth | >5MHz |
Special functions |
|
Product Ohmic drop compensation | 2V/current range, 16bit resolution |
Peripheral connections |
|
Product Shared analog input/output | User selectable analog input: ±10V, 16bit resolution, or analog output: ±10V, 16bit resolution, bandwidth 40kHz |
Envionment |
|
Product Power requirements | 100-240V, 45-65Hz, 4VA via 5V supply (included) |
Product Interfacing | USB |
Product Size (w x d x h) | 10 x 19 x 2.5cm |
Product Weight | 500g |
Product PC requirements | Windows 7/8/10, with free USB port |