The glassy carbon electrode (GC) is the most used working electrode in voltammetry. Glassy carbon is hard, gas-impermeable and chemically inert, with a wide potential window and an easy-to-polish surface. It is the basis for modified electrodes, sensors and electrocatalysis studies.
Why use glassy carbon
- Wide potential window: about −1.0 V to +1.2 V vs. Ag/AgCl in aqueous medium, depending on pH and electrolyte.
- Low background current: it favors the detection of low concentrations.
- Chemical inertness: resists acids, bases and organic solvents.
- Renewable surface: polishing returns a clean, reproducible surface.
Applications
- Cyclic and differential pulse voltammetry for the characterization of compounds.
- Sensors and biosensors with modified electrodes (nanotubes, graphene, enzymes, polymers).
- Electrocatalysis with catalyst paints deposited on the disc, including on a rotary disk electrode.
- Analysis of drugs, food and environmental contaminants.
Polishing and cleaning
- Polishing in "eight" movement on a polishing cloth with alumina suspension, from the largest grain to the smallest (for example, 1 µm, 0.3 µm and 0.05 µm).
- Rinse with ultrapure water between each step.
- Make a quick ultrasound bath in water or ethanol to remove alumina residues.
- Checking the surface by cyclic voltammetry in potassium ferricyanide: well-defined peaks and separation close to 60–70 mV indicate a good surface.
How to choose
- Disc diameter: 3 mm is the default; 1 to 2 mm for small volumes; 5 mm for rotary disc electrodes.
- Body: PEEK resists organic solvents; PTFE (Teflon) and Kel-F are classic options.
- Connection: pin or thread compatible with the potentiostat cable or rotator.
- Accessories: polishing kit with alumina and cloths.
View the models in work electrodes, os reference electrodes e o reference electrodes Guide.
FAQ
How often should I polish?
Before each series of measurements, or whenever the peaks lose definition. Modified electrodes are polished only when removing the film.
Vitreous or platinum carbon?
The glassy carbon has a larger cathodic window and a lower reactivity; Platinum is preferred for hydrogen and oxygen adsorption studies and for many electrocatalysis processes.