Ideal for research applications that require precise control, extensive experimental flexibility and support for impedance-based techniques.
_edited_edited.png)




Potentiostat/single channel with FRA
Current Range: ±1 A max.
Voltage Scan Range: ±10 V
EIS frequency range: 10 µHz - 2 MHz
Compatible with Windows, macOS and Ubuntu through the Squidstat and API (Application Programming Interface).
Integrated memory (16 GB) automatically backs up all channel data for future recovery.
Multiple units can work simultaneously with other Squidstat models.
Made in the USA

Galvanostatic electrochemical impedance spectroscopy (GIs)
Electrochemical impedance spectroscopy Potentiostatic (PEIS)
Mott-Schottky (single frequency and multifrequency)
Battery Capacity
Load/Disload (C constant / V constant)
Load/discharge (Constant Resistance / Power constant)
Chronoampperometry
Chronopotentiometry
Constant power
Constant resistance
Cyclic Voltametry (VC)
Differential Pulse Voltammetry (DPV)
Galvanostatic Intermittent Titration Technique (GITT)
Intermittent potentiostatic titration technique (PITT)
Amperometry Linear Scan
Linear Scan (LSV) Voltammetry
Normal Pulse Voltammetry
Open Circuit Potential Monitoring
Rotary Ring Disk Electrode (RRDE)
Square Wave Voltammetry
Voltammetry in Ladder
Direct Current Experiments
AC Experiments
DC Parameters
Working Electrode (V)
Working Electrode (MV)
DC Working Electrode (V)
DC Working Electrode (MV)
Counterelectrode (V)
Counterelectrode (MV)
Current (A)
Current (MA)
DC Current (A)
DC Current (MA)
Working Electrode vs. NHE (V)
Current Density (A/M2)
Current Density (A/cm²)
Power (W)
Accumulated Load (MAH)
Accumulated Load (C)
Capacity (Mah)
Energy (WH)
Load status (%)
Discharge depth (%)
dQ/dV (mAh/V)
1 / √(Time) (1/√S)
Resistance (OHM)
Number of cycles
Parameters CA
Frequency (Hz)
Phase (Grades)
|Z| (ohms)
Z' (ohms)
-z'' (ohms)
Z'' (ohms)
Amplitude (A)
Amplitude (V)
Amplitude (MV)
Duration of each Frequency (s)
Number of Cycles
THD (%)
Serial ((1/C2)(1/µF2))
Serial ((1/C)(1/µF))
Parallel ((1/C2)(1/µF2))
Parallel ((1/C)(1/µF))
Additional Parameters
PC Date/Time Stamp
UTC time(s)

|
Potentiostatic Input/Output
|
Values
|
|---|---|
|
Compliance Voltage
|
±12 V
|
|
Voltage Scan Range
|
±10 V
|
|
Number of Voltage Ranges
|
4 Ranges, [±10 V, ±5 V, ±1 V, ±0.5V]
|
|
Applied Voltage Accuracy
|
0.1% of the reference value, 1 mV min
|
|
Applied Voltage Resolution
|
0.003% (30.52 ppm) of the range, 15 μV min
|
|
Measured voltage accuracy
|
0.1% of the measured value, 1 mV min
|
|
Voltage resolution measured
|
0.003% (30.52 ppm) of the range, 15 μV min
|
|
Time resolution
|
10 ns
|
|
IR compensation limits
|
23.5% of the operating range, 2.35 V max.
|
|
Galvanostatic input/output
|
Values
|
|---|---|
|
Current range
|
±1 A
|
|
Number of current intervals
|
8 ranges, [100 nA A 1 A]
|
|
Applied Current Accuracy
|
0.2% of the range, 1
|
|
Applied Current Resolution
|
0.003% (30.52 ppm) of the range, 3 Pa min
|
|
Measurement accuracy
|
0.1% of the range, 100 Pa min
|
|
Measured current resolution
|
0.003% (30.52 ppm) of the band, 3 Pa min
|
|
General Specifications
|
Values
|
|---|---|
|
Channels per Unit
|
1
|
|
Operating Modes
|
Pot / Gal / ZRA / FRA
|
|
Cellular Connections
|
5 Electrodes
|
|
ADC/DAC resolution
|
16-bit resolution
|
|
Variation rate (1 kΩ load)
|
1 MV/s (-1 V a +1 V)
|
|
Minimum DC data recording interval
|
200 µs
|
|
Shorter DC data sampling interval
|
7,5 µs
|
|
Mode Floating/Terrestrial
|
Both (changeable)
|
|
Input Impedance
|
10 tera-ω (Typical)
|
|
Input Polarization Current
|
10 PA (Typical)
|
|
Data Backup Memory
|
16 GB integrated
|
|
Dimensions and Weight
|
24 cm long x 17 cm wide x 6 cm high, 1 kg
|
|
Channel cable length
|
1 m
|
|
Computer interface
|
1 USB per unit
|
|
Operating temperature range
|
0 a 40 ℃
|
|
Minimum/maximum scanning rate
|
15 µV/s minimum, 25 V/s maximum with 10 mV sampling
|
|
Smaller pulse width
|
375 µs
|
|
Lower step of potential for scanning voltammetry
|
0.003% (30.52 ppm) of the range, 15 μV min
|
|
Square wave voltammetry Maximum frequency
|
3.5 kHz (-1 V to +1 V)
|
|
Power supply requirements
|
100 – 240 VAC, 50 – 60 Hz
|
|
Impedance analyzer
|
Values
|
|---|---|
|
EIS GALVANOSTACTIC's amplitude resolution
|
0.000012% (0.12 ppm) of the range
|
|
Breadth is galvanostatic
|
≤ 100% of the range, 1 to max.
|
|
EIS Potentiostatic Span Resolution
|
0.000012% (0.12 ppm) of the setpoint/measurement
|
|
Amplitude EIS potentiostatic
|
≤ 100% of the range, 10 V max.
|
|
EIS amplitude resolution
|
24-bit resolution
|
|
AC frequency resolution
|
18 bits, 0.0004% (3.81 ppm), 3 µHz minimum
|
|
AC frequency accuracy
|
0.005% (50 ppm) or better
|
|
EIS FREQUENCY RANGE
|
10 μHz to 2 MHz
|