Ideal for research environments that require high power management capacity, precise control and advanced analytics techniques in a compact system.

Potentiostat/Galvanostat Single with FRA
Current range: ±5 to max.
Voltage Scan Range: ±10 V
EIS frequency range: 10 µHz - 2 MHz
Compatible with Windows, macOS and Ubuntu through the Squidstat and API (Application Programming Interface).
Integrated memory (16 GB) automatically backs up all channel data for future recovery.
Multiple units can work simultaneously with other Squidstat models.
Made in the USA





Galvanostatic electrochemical impedance spectroscopy (GIs)
Spectroscopy of Potentiostatic electrochemical impedance (PEIS)
Mott-Schottky (single frequency and multifrequency)
Battery Capacity
Load/Disload (C constant / V constant)
Load/Disload (Constant Resistance / Constant Power)
Chronoamperometry
Chronopotentiometry
Constant Power
Constant Resistance
Cyclic Voltametry (VC)
Differential Pulse Voltammetry (DPV)
Galvanostatic Intermittent Titration Technique (GITT)
Intermittent potentiostatic titration technique (PITT)
Amperometry Linear Scan
Linear Scan (LSV) Voltammetry
Normal Pulse Voltammetry
Open Circuit Potential Monitoring
Rotary Ring Disk Electrode (RRDE)
Square Wave Voltammetry
Voltammetry in Ladder
Direct Current Experiments
AC Experiments
DC Parameters
Working Electrode (V)
Working Electrode (MV)
DC Working Electrode (V)
DC Working Electrode (MV)
Counterelectrode (V)
Counterelectrode (MV)
Current (A)
Current (MA)
DC Current (A)
DC Current (MA)
Working Electrode vs. NHE (V)
Current Density (A/M2)
Current Density (A/cm²)
Power (W)
Accumulated Load (MAH)
Accumulated Load (C)
Capacity (Mah)
Energy (WH)
Load status (%)
Discharge depth (%)
dQ/dV (mAh/V)
1 / √(Time) (1/√S)
Resistance (OHM)
Number of cycles
Parameters CA
Frequency (Hz)
Phase (Grades)
|Z| (ohms)
Z' (ohms)
-z'' (ohms)
Z'' (ohms)
Amplitude (A)
Amplitude (V)
Amplitude (MV)
Duration of each Frequency (s)
Number of Cycles
THD (%)
Serial ((1/C2)(1/µF2))
Serial ((1/C)(1/µF))
Parallel ((1/C2)(1/µF2))
Parallel ((1/C)(1/µF))
Additional Parameters
Temperature (Grauses, °C)
PC Date/Time Stamp
Time UTC (S)

|
Potentiostatic Input/Output
|
Values
|
|---|---|
|
Compliance Voltage
|
±11,5 V min
|
|
Voltage Scan Range
|
±10 V
|
|
Number of Voltage Ranges
|
4 Ranges, [±10 V, ±5 V, ±1 V, ±0.5V]
|
|
Applied Voltage Accuracy
|
0.1% of the reference value, 1 mV min
|
|
Applied Voltage Resolution
|
0.003% (30.52 ppm) of the range, 15 μV min
|
|
Measured voltage accuracy
|
0.1% of the measured value, 1 mV min
|
|
Voltage resolution measured
|
0.003% (30.52 ppm) of the range, 15 μV min
|
|
Time resolution
|
10 ns
|
|
IR compensation limits
|
12% of the operating range, 1.2 V max.
|
|
Input/Output Galvanostatic
|
Values
|
|---|---|
|
Current Range
|
±5 A
|
|
Number of Current Intervals
|
8 Ranges, [1 μA to 5 A]
|
|
Applied Current Accuracy
|
0.1% of the range, 1 in the min
|
|
Applied current resolution
|
0.003% (30.52 ppm) of the range, 30 Pa min
|
|
Accuracy of current measured
|
0.1% of the range, 1 in the min
|
|
Measured current resolution
|
0.003% (30.52 ppm) of the band, 30 Pa min
|
|
General Specifications
|
Values
|
|---|---|
|
Channels per unit
|
1
|
|
Operating modes
|
Pot / Gal / ZRA / FRA
|
|
Cellular connections
|
5 Electrodes
|
|
ADC/DAC resolution
|
16-bit resolution
|
|
Variation rate (1 kΩ load)
|
1 MV/s (-1 V a +1 V)
|
|
Minimum DC data recording interval
|
200 µs
|
|
Shorter DC data sampling interval
|
7,5 µs
|
|
Mode Floating/Terrestrial
|
Both (changeable)
|
|
Input Impedance
|
10 tera-ω (Typical)
|
|
Input Polarization Current
|
10 PA (Typical)
|
|
Data Backup Memory
|
16 GB integrated
|
|
Dimensions and Weight
|
24 cm long x 17 cm wide x 13 cm high, 3.5 kg
|
|
Channel cable length
|
1 m
|
|
Computer interface
|
1 USB per unit
|
|
Operating temperature range
|
0 a 40 ℃
|
|
Minimal/maximum scan rate
|
15 µV/s Minimum, 25V/s maximum with sampling of 10 mV
|
|
Lower pulse width
|
375 µs
|
|
Lower potential step for scanning voltammetry
|
0.003% (30.52 ppm) of the range, 15 μV min
|
|
Square wave voltammetry Maximum frequency
|
2 kHz (-1 V to +1 V)
|
|
Power supply requirements
|
100 – 240 VAC, 50 – 60 Hz
|
|
Impedance Analyzer
|
Values
|
|---|---|
|
EIS Frequency Range
|
10 μHz to 2 MHz
|
|
Accuracy of the frequency AC
|
0.005% (50 ppm) or better
|
|
AC frequency resolution
|
18 bits, 0.0004% (3.81 ppm), 3 µHz minimum
|
|
EIS amplitude resolution
|
24-bit resolution
|
|
EIS Potentiostatic Range
|
≤ 100% of the range, 10 V max.
|
|
EIS Potentiostatic EIS amplitude resolution
|
600 nV
|
|
EIS Galvanostactic Amplitude
|
≤ 100% of the range, 5 a max.
|
|
EIS galvanostatic EIS galvanostatic resolution
|
0.0006% (6 ppm) of the track
|
Highlights of Squidstat Penta:
Each Squidstat Penta includes a 100 cm working electrode cable, a 100 cm counter electrode cable, a 100 cmSense cable with reference connections, working directions, direction, a USB cable and a power cable for Connect to an electrical outlet. The Squidstat software User interface and unlimited remote technical support are included in the purchase.
|
|
|
|
Channels per unit |
1 |
|
Operating modes |
Pot / Gal / ZRA / FRA |
|
Cellular connections |
2, 3, 4 or 5 electrodes |
|
ADC/DAC resolution |
16-bit resolution |
|
Rate Swing (without load) |
5MV/s |
|
Maximum sampling speed |
100,000 samples/s |
|
Floating and/or grounding |
Both (changeable) |
|
Input impedance |
> 10 Tera-ohm |
|
Input Polarization Current |
< 1 pA |
|
Data backup memory |
16 GB |
|
Physical dimensions |
23 cm x 17 cm x 13 cm |
|
Channel cable length |
100 cm |
|
Computer interface |
1 USB per unit |
|
Power supply requirements |
100 – 240 VAC 50 – 60 Hz |
|
Compliance voltage |
±10 V |
|
Range of Voltage Sweep |
±10 V |
|
Number of voltage ranges |
4 ranges, (0,5V, 1V, 5V, 10V) |
|
Applied Potential Precision |
0.1% of the setpoint, 1 mV min |
|
Applied Potential Resolution |
0.003% of the range, 15 μV min |
|
Measured potential accuracy |
0.1% reading, 1 mV min |
|
Potential resolution measured |
0.003% range, 15 μV min |
|
Maximum current |
±5 A |
|
Current intervals |
8 ranges, (1 μA a 5 A) |
|
Current applied accuracy |
0.1% of the range, 1 nA min |
|
Current resolution applied |
0.003% of range, 30 pA min |
|
Current Accuracy Measure |
0.1% of the range, 1 nA min |
|
Current resolution Measure |
0.003% of range, 30 pA min |
|
Frequency range EIS |
10 μHz to 1 MHz |
|
AC frequency accuracy |
0.005% or better |
|
AC frequency resolution |
0.0004%, 3 µHz min |
|
Potentiostatic amplitude |
≤ 100% of the range, 10 V máx. |
|
potentiostatic Amplitude Resolution |
25 µV @ 1mV 250 µV @ 100mV 2,5 mV @ 10 V |
|
Galvanostatic Amplitude |
≤ 100% of Range, 5 A máx. |
|
galvanostatic Amplitude Resolution |
0.025% of applied to 100% of the range |
LabSolutions is the exclusive representative of Admiral Instruments in Brazil.