Ideal for advanced research and industrial applications where reliability at high current is essential.

Single-channel potentiostat/galvanostat with FRA
Current range: ±20 A max
Voltage scan range: ±6 V
EIS frequency range: 10 µHz - 2 MHz
Compatible with Windows, MacOS, and Ubuntu via the Squidstat User Interface software and Application Programming Interface
Onboard memory (16 GB) automatically backs up all channel data for future recovery
Multiple units can run simultaneously with other Squidstat models
Made in USA





Galvanostatic Electrochemical Impedance Spectroscopy (GEIS)
Potentiostatic Electrochemical Impedance Spectroscopy (PEIS)
Mott-Schottky (single-frequency and multi-frequency)
Battery Capacity
Charge / Discharge (Constant C / Constant V)
Charge / Discharge (Constant R / Constant P)
Chronoamperometry
Chronopotentiometry
Constant Power
Constant Resistance
Cyclic Voltammetry (CV)
Differential Pulse Voltammetry (DPV)
Galvanostatic Intermittent Titration Technique (GITT)
Potentiostatic Intermittent Titration Technique (PITT)
Linear Sweep Amperometry
Linear Sweep Voltammetry (LSV)
Normal Pulse Voltammetry
Open Circuit Potential Monitoring
Rotating Ring Disk Electrode (RRDE)
Square Wave Voltammetry
Staircase Voltammetry
DC Experiments
AC Experiments
DC Parameters
Working Electrode (V)
Working Electrode (mV)
DC Working Electrode (V)
DC Working Electrode (mV)
Counter Electrode (V)
Counter Electrode (mV)
Current (A)
Current (mA)
DC Current (A)
DC Current (mA)
Working Electrode vs. NHE (V)
Current Density (A/m2)
Current Density (A/cm2)
Power (W)
Cumulative Charge (mAh)
Cumulative Charge (C)
Capacity (mAh)
Energy (Wh)
State of Charge (%)
Depth of Discharge (%)
dQ/dV (mAh/V)
1 / Sqrt(time) (1/√s)
Resistance (Ohm)
No of Cycles
AC Parameters
Frequency (Hz)
Phase (deg)
|Z| (Ohms)
Z' (Ohms)
-Z'' (Ohms)
Z'' (Ohms)
Amplitude (A)
Amplitude (V)
Amplitude (mV)
Duration of Each Frequency (s)
No of Cycles
THD (%)
Serial ((1/C2)(1/µF2))
Serial ((1/C)(1/µF))
Parallel ((1/C2)(1/µF2))
Parallel ((1/C)(1/µF))
Additional Parameters
Temperature (Deg, °C)
PC Timestamp
UTC Time (s)

|
Potentiostatic Input/Output
|
Values
|
|---|---|
|
Compliance Voltage
|
±6 V
|
|
Voltage Scan Range
|
±6 V
|
|
Number of Voltage Ranges
|
4 ranges, [±6 V, ±5 V, ±1 V, ±0.5 V]
|
|
Applied Voltage Accuracy
|
0.1% of setpoint, 1 mV min
|
|
Applied Voltage Resolution
|
0.003% (30.52 ppm) of range, 15 μV min
|
|
Measured Voltage Accuracy
|
0.1% of measured, 1 mV min
|
|
Measured Voltage Resolution
|
0.003% (30.52 ppm) of range, 15 μV min
|
|
Time Resolution
|
10 ns
|
|
iR Compensation Limits
|
1 V
|
|
Galvanostatic Input/Output
|
Values
|
|---|---|
|
Current Range
|
±20 A
|
|
Number of Current Ranges
|
9 ranges, [1 μA to 20 A]
|
|
Applied Current Accuracy
|
0.1% of range ≤ 10 A, 0.2% > 10 A
|
|
Applied Current Resolution
|
0.003% (30.52 ppm) of range, 30 pA min
|
|
Measured Current Accuracy
|
0.1% of range, 1 nA min
|
|
Measured Current Resolution
|
0.003% (30.52 ppm) of range, 30 pA min
|
|
General Specifications
|
Values
|
|---|---|
|
Channels Per Unit
|
1
|
|
Operating Modes
|
Pot / Gal / ZRA / FRA
|
|
Cell Connections
|
5 electrode
|
|
ADC/DAC Resolution
|
16-bit resolution
|
|
Slew Rate (1 kΩ load)
|
1 MV/s (-1 V to +1 V)
|
|
Shortest DC data recording interval
|
200 µs
|
|
Shortest DC data hardware sampling interval
|
7.5 µs
|
|
Floating/Ground Mode
|
Both (Switchable)
|
|
Input Impedance
|
10 Tera-Ω (typical)
|
|
Input Bias Current
|
6 pA (typical)
|
|
Data Backup Memory
|
16 GB onboard
|
|
Dimension and Weight
|
24 L x 17 W x 13 H cm, 3.5 kg
|
|
Channel Cable Length
|
1 m
|
|
Computer Interface
|
1 USB per unit
|
|
Temperature Operating Range
|
0 to 40 ℃
|
|
Min/Max Scan Rate
|
15 µV/s minimum, 25 V/s maximum with 10 mV sampling
|
|
Smallest Pulse Width
|
375 µs
|
|
Smallest Potential Step for Sweep Voltammetry
|
0.003% (30.52 ppm) of range, 15 μV min
|
|
Square Wave Voltammetry Max Frequency
|
2 kHz (-1 V to +1 V)
|
|
Power Supply Requirements
|
100 – 240 VAC, 50 – 60 Hz
|
|
Impedance Analyzer
|
Values
|
|---|---|
|
EIS Frequency Range
|
10 μHz to 2 MHz
|
|
AC Frequency Accuracy
|
0.005% (50 ppm) or better
|
|
AC Frequency Resolution
|
18-bit, 0.0004% (3.81 ppm), 3 µHz min
|
|
EIS Amplitude Resolution
|
24-bit resolution
|
|
Potentiostatic EIS Amplitude
|
≤ 100% of range, 6 V max
|
|
Potentiostatic EIS Amplitude Resolution
|
0.000012% (0.12 ppm) of setpoint/measured
|
|
Galvanostatic EIS Amplitude
|
≤ 100% of range, 20 A max
|
|
Galvanostatic EIS Amplitude Resolution
|
0.000012% (0.12 ppm) of range
|
±20 A Potentiostat with EIS up to 1 MHz and 1% / 1° Accuracy at 0.01 mΩ
Squidstat Venta Highlights:
Each Squidstat Venta includes one 90 cm Working Electrode cable, one 90 cm Counter Electrode cable, one sense cable with Reference, Working Sense, Counter Sense connections, one USB cable, and one power cable to connect to a power outlet. The Squidstat User Interface software and unlimited remote technical support are included with purchase.
|
Price Per Unit ($USD) |
$16,900 |
|
Channels Per Unit |
1 |
|
Operating Modes |
Pot / Gal / ZRA / FRA |
|
Cell Connections |
2, 3, 4, or 5 electrode |
|
ADC/DAC Resolution |
16 bit resolution |
|
Slew Rate (No Load) |
5 MV/s |
|
Maximum Sampling Speed |
100,000 samples/s |
|
Floating and/or Earth-grounding |
Both (Switchable) |
|
Input Impedance |
> 10 Tera-ohm |
|
Input Bias Current |
< 1 pA |
|
Data Backup Memory (Onboard) |
16 GB |
|
Physical Dimensions |
23 cm x 17 cm x 13 cm |
|
Channel Cable Length |
90 cm |
|
Computer Interface |
1 USB per unit |
|
Power Supply Requirements |
100 – 240 VAC, 50 – 60 Hz |
|
Compliance Voltage |
±6 V per channel |
|
Scan Range (Controlled Voltage) |
1 range, ±6 V |
|
Applied Potential Accuracy |
0.1% of setpoint, 2 mV max accuracy |
|
Applied Potential Resolution |
180 µV |
|
Measured Potential Accuracy |
0.1% of setpoint, 1 mV max accuracy |
|
Measured Potential Resolution |
180 µV |
|
Maximum Current |
±20 A per channel |
|
Current Ranges |
9 ranges (1 μA to 20 A) |
|
Applied Current Accuracy |
0.1% of range, 1 nA max |
|
Applied Current Resolution |
0.003% of range, 30 pA max |
|
Measured Current Accuracy |
0.1% of range, 1 nA max |
|
Measured Current Resolution |
0.003% of range, 30 pA max |
|
EIS Frequency Range |
10 μHz to 1 MHz |
|
AC Frequency Accuracy |
0.005% or better |
|
AC Frequency Resolution |
0.0004%, 3 µHz max |
|
Potentiostatic Amplitude |
≤ 10% of range, 1 V max |
|
Pot. Amplitude Resolution |
240 µV, 12 bit resolution |
|
Galvanostatic Amplitude |
≤ 10% of range, 2 A max |
|
Gal. Amplitude Resolution |
0.002% of range, 24 pA max |