The ISX-5 combines up to eight EIS independent measurement modules in a single chassis.
Do you need many stable positions, but not parallel acquisitions? A ISX-3 with MUX may be the simplest solution.

INDEPENDENT EIS MODULESDedicated measurement modules acquire multiple impedance paths in parallel, instead of switching a single analyzer between inputs.
CORRESPONDING TIME SIGNALLow-level timing is aligned between modules for simultaneous events and consistent phase measurements across modules.
BROADBAND EIS PER MODULEComplete impedance spectroscopy in all parallel measurement paths.
UP TO 512 CONNECTED POSITIONSAdd MUX32/64 after selected modules when the system also needs many sequential positions.
A multiplexer allows a single analyzer to reach multiple positions by measuring them sequentially. This is usually the simplest and most efficient way when testing devices (DUTs) remain stable during scanning.
ISX-5 uses independent measuring modules to acquire several impedance paths in parallel. This is important when samples change during acquisition, when multiple responses need to be compared for the same event or when simultaneous execution of multiple complete measurements significantly reduces the time of the experiment.
Buyer's Rule — If sequential measurements describe the experiment correctly, use multiplexing. If multiple impedance measurements actually need to be performed simultaneously, use EIS in parallel.
Channel 1 -> Channel 2 -> Channel 3 -> Channel 4.
Ideal when multiple positions can be measured one after another.
Channel 1 =====\ Channel 2 ======= channel 3 ======= channel 4 =====
Better when multiple impedance paths need to be acquired in the same time interval.
ISX-5 is not an analyzer with several simultaneous inputs. Each parallel path has its own full impedance measurement module, with low-level synchronization between modules.
This architecture offers more than several simultaneous measurements: it allows for synchronized operation, independent configuration of channels, measurements between modules with phase relation preservation and parallel throughput.
Coordinate multiple modules around the same measurement sequence, time window, or experimental event.
Use different frequency programs, excitation levels, electrode settings or measurement settings in different modules.
Because the timing of the modules is strictly synchronized, the excitation of one module can be measured by another, preserving the phase relationship required for the complex impedance.
Acquire multiple EIS complete measurements simultaneously, even when strict concurrency is not the only system requirement.
Module 1 -> DUT 1 · Module 2 -> DUT 2 · Module 3 -> DUT 3 · ... · Module 8 -> DUT 8.
Each module provides an independent EIS path.
For applications that require more than eight parallel EIS modules, multichassis architectures can be designed specifically for each application.
Configure up to eight independent measurement modules in a single chassis. Run the modules synchronously, use independent measurement settings when necessary, and add multiplexing behind selected modules where the system also needs sequential positions.
Up to 8 in a single chassis
Each module provides its own full EIS measurement path.
Accurate low-level signal synchronization
Align simultaneous events between channels and preserve the phase relationship for measurements between modules.
Synchronous or independent operation; individual settings per module.
Run the modules together or use different measurement programs and excitation settings as needed.
MUX32 / MUX64 behind selected modules; Up to 8 multiplexers
Combine parallel acquisition with sequential expansion; Up to 512 positions connected with 8 x MUX64.
Application-specific multichassis architectures
For projects that need more than 8 parallel paths.
Hardware synchronization input/output options and external control
Coordinate the acquisition of EIS data between modules, chassis and external experimental events.
100 MHz – 10 MHz
EIS broadband per measurement module; Extended frequency settings by technical analysis.
1 ... 2048 by sweep
Detailed programs of linear, logarithmic and block frequency.
mOhm ... GOhm specified
Use the precision contour for the actual impedance/frequency operation region.
0.1% specified
The wide central reference region reaches 0.01% under documented conditions.
0,1°
Phase precision specified for EIS.
1 mV... 1 V peak
Controlled tension excitation.
10 µA ... peak of 10 mA
Controlled current excitation.
2, 3 and 4 electrodes
Adapt the topology to the device under test (DUT) and the effects of the connection.
±1 V
Combined configuration and excitation/polarization limits apply.
USB and ethernet
PC operation and programmatic control.
ExtensionPort connector options/front panel
User interfaces, adapters and sensors specific for each application, when necessary.
BNC, SMA, MCX, Banana, DSUB or Custom connectors, depending on the configuration.
Select the physical interface around the device under test (DUT) and the channel architecture.
The useful performance of each measurement module ISX-5 depends on the combination of frequency, impedance and the complete connection path. Therefore, the precision curve is more significant than a single ideal value.
In hybrid systems, multiplexers, cables, fixing devices and application-specific interfaces become part of the actual measurement chain and should be reviewed when the operating point is demanding.
Show the precision contour of the released ISX and highlight the region relevant to the parallel experiment.
Use a dynamic test device (DUT) or an example between modules to demonstrate event time matching and consistently in phase acquisition.
ISX-5 performs better when it is necessary to acquire several samples, sensors or measurement locations simultaneously, compare them through the same event, or measure them with consistent phase relationships between the modules.
Use parallel electrochemical impedance spectroscopy (EIS) for cultures, tissues, biochips or physiological models where different regions or conditions respond during the same experiment.
Purchase multiple sensor channels in parallel when the response time or channel comparison is part of the result.
Compare samples during temperature changes, fluid dynamic events, exposure to chemicals, mechanical stimulation or other controlled conditions.
Use parallel modules when different regions, electrode groups or measurement paths need to be compared without delay in the order of scanning.
Use strictly synchronized modules when excitation and detection cover different modules and the complex impedance phase needs to remain significant.
Combine parallel acquisition with external control, synchronization and software integration when impedance measurement is only a part of a larger experiment.




Each measurement module ISX-5 can also be combined with a MUX32 or MUX64. This allows the system to maintain independent parallel measurement paths where they are needed, adding sequential positions after each module.
With eight modules equipped with MUX64, a ISX-5 chassis can address up to 512 connected positions, maintaining up to eight independent parallel EIS paths. Larger multiplexed configurations such as 128 ports per module are possible as application-specific systems.
Each module connects directly to a device under test (DUT) or measuring path.
Four modules remain in parallel; The positions behind each MUX are measured sequentially.
Eight modules remain in parallel; Each module can scan up to 64 positions sequentially.
Start with the impedance paths that really need parallel measurement. Then set any multiplexed positions, measurement settings, interfaces and synchronization requirements.
1
2
3
4
5
6
7
8
The default is BNC.
BNC
SMA
MCX
Banana connector
DSUB
Other/Custom
10 MHz
40 MHz
Io door
Additional connector with digital I/O, serial interfaces and NTC temperature sensors for connecting external devices, intelligent sensors, etc.
Medical power supply
Insulation class II, in accordance with the EN60601-1 standard.
Expand your configuration by adding multiplexers to the instrument. Note that each multiplexer needs a primary measurement channel.
32-channel multiplexer, any-for-any-channel
Highly flexible multiplexer with 32 ports, allowing any combination of impedance measurement ports with 2, 3 or 4 electrodes. Connector type: MDR40
64-channel multiplexer, any-for-any-channel
Highly flexible multiplexer with 64 ports, allowing any combination of impedance measurement ports with 2, 3 or 4 electrodes. Connector type: 2 x MDR40
Second 4 electrode configuration channel
second measuring channel multiplexed with the same connector types as the primary channel
AC coupling Interface
Y2 capacitors in accordance with IEC 60384-14:2013 with high levels of insulation. With the internal circuit operating in the SELV domain, they serve MOPP ratings of 2.
Medical grade HS USB Insulator
Medical grade ethernet insulator
BNC - BNC 1 m
BNC cables with 1 m Kelvin clip
Cable compatible with the internal multiplexer module connector, with the termination of your choice (pins, BNC, SMA, etc.).
BNC - 4 mm, 1 m banana plug
BNC - MCX 0.8m
1 m MUX cable set
RG178; MDR - Banana Plug (+ Alligator Claws)
Base unit
ISX-5
Number of measurement channels
—
Front panel connectors
—
Channel
—
Additional functionality
—
Integrated multiplexers
—
Accessories
—
Measuring cable sets
—
Our application engineers will analyze your configuration.
Talk to an ExpertConfigure spectra, module definitions and electrode topology in the ISX software. For more complex systems, define multiplexed paths, coordinate the acquisition and integrate the analyzer via USB, Ethernet, synchronization and I/O routes.
Configure frequency, excitation, accuracy and topology programs in the independent measurement modules.
Coordinate channels for synchronization or configure different measurement programs as needed for the devices under test (DUTs).
Configure C, R, WS and W functions in direct and multiplexed paths.
Use command control, synchronization and I/O when the EIS acquisition is part of a larger automated configuration.
Use this tool when the electrode samples or positions can be measured one after the other and the parallel acquisition is not required.
Use when a single impedance path is sufficient and the configuration does not require parallel channels.
Use when multi-electrode spatial acquisition and detailed spectroscopy are appropriate together.
Use this solution when routing 64-door sensors, electrochemical modes and a compact hybrid architecture are best suited.
Use ISX-5 to determine how many impedance paths need parallel procurement, which can be multiplexed, and which timing and measurement settings are important.
Fix channel mapping, the interface of the device under test (DUT), the synchronization, the software workflow and the automation around the proven measurement.
Use validated requirements to define the scope of the dedicated implementation in electronics, firmware, software, mechanics and production.
ISX-5 is the truly parallel multi-channel impedance analyzer Sciospec. It combines up to eight independent EIS measurement modules in a single chassis, allowing the acquisition of multiple impedance paths in parallel rather than sequentially.
Let us know which paths should be run in parallel, the speed of change of samples, whether measurement between modules is necessary and how many additional positions can be sequential. We will recommend the ISX-5, the hybrid ISX-5 + MUX, the multiple chassis analysis, or the ISX-3 + MUX.
Select the number of parallel modules, MUX options, interfaces and synchronization requirements and send the configuration for review and budget.
Use this method when EIS parallel becomes part of a recurring test system, reader or OEM product.