The reference electrode provides a stable and known potential against which the potentiostat measures and controls the potential of the working electrode. An unstable or poorly chosen reference electrode displaces peaks, distorts voltammograms and compromises reproducibility. This guide compares the most used types and explains how to choose and conserve each one.
Main types
- Ag/AgCl: silver thread coated with silver chloride in KCl solution. With 3 M KCl, it is about +0.21 V vs. EPH; With saturated KCl, about +0.197 V. It is the most used because it is robust and free of mercury.
- Saturated calomel (SCE): Hg/Hg₂Cl₂ In saturated KCl, about +0.241 V vs. EPH. Very stable, but contains mercury and should not exceed about 50 °C.
- Hg/HgO: suitable for alkaline media (NaOH or KOH), such as in water electrolysis and alkaline batteries.
- Hg/Hg₂SO₄: for media with sulfate and when chloride should be avoided.
- Ag/Ag⁺ Non-aqueous: for organic solvents such as acetonitrile; It is customary to report the potential against the Ferrocene/Ferrocene pair.
- Reversible hydrogen electrode (ERH): reference in the solution itself, widely used in electrocatalysis and fuel cells.
How to choose
- Measurement medium: acid or neutral with chloride → Ag/AgCl or SCE; Alkaline → Hg/HgO; Organic → Ag/Ag⁺.
- Chloride contamination: if the chloride hinders the measurement, use double saline or Hg/Hg₂SO₄.
- Temperature: check usage range; The SCE loses stability above about 50 °C.
- Format: diameter and length compatible with the electrochemical cell; Mini versions for microcells.
- Junction: fry ceramic, porous glass or glove junction, depending on the electrolyte flow and the risk of clogging.
CARE AND MAINTENANCE
- Store the electrode with the tip immersed in the same filling solution (eg, 3 m KCl), never in pure water.
- Complete the internal solution when the level drops and avoid air bubbles.
- Check the potential periodically against a new reference electrode: Differences above some MVs indicate maintenance requirements.
- If the fry clogs or darkens, replace it or replace the electrode.
View the models in reference electrodes, as electrochemical Cells e os working Electrodes.
FAQ
How to convert potentials between Ag/AgCl and SCE?
Add the difference between the potentials of the references: A value against Ag/AgCl (KCl 3M) is about 0.03 V more negative when expressed against SCE. Always indicate the reference and concentration of KCl.
Can I store the dry electrode?
For a short time, with the cover and the internal solution. For frequent use, keep the tip immersed in the filling solution.
Why is the potential oscillating?
The most common causes are air bubble at the tip, clogged joint or contaminated internal solution.