Electrochemical impedance spectroscopy (EIS) measures the response of an electrochemical system to a small alternating signal over a wide range of frequencies. With it, it is possible to separate processes that take place at different time scales, such as load transfer, diffusion and electrolyte resistance. It is used in batteries, fuel cells, corrosion, coatings, sensors and biosensors.
How the measurement works
The potentiostat applies a small sinusoidal (potentiostatic mode) or current (galvanostatic mode) disturbance typically from 5 to 10 mV over the operating point, and measures the response. The ratio between the applied signal and the response, with its lag, is the impedance. The scan usually goes from some MHz or kHz to MHz.
How to read the results
- Nyquist diagram: imaginary part against the real part of the impedance. semicircles indicate load transfer processes; Lines at 45° indicate diffusion (Warburg).
- Bode diagram: module and phase as a function of frequency, useful for identifying the time constants of the system.
- Equivalent circuits: resistors, capacitors and constant phase elements (CPE) model the physical processes and allow you to extract parameters such as load transfer resistance and double-layer capacitance.
Applications
- Batteries: internal resistance, health status (SoH) and aging.
- Corrosion and coatings: efficiency of inhibitors and degradation of paintings.
- Sensors and Biosensors: detection by impedance variation on the electrode surface.
- Fuel cells and electrolyzers: ohmic, activation and mass transport losses.
- Biology and materials: characterization of cells, tissues and materials, and impedance tomography.
How to choose a potentiostat for EIS
- Frequency range: up to 100 kHz meets most electrochemical applications; Up to 1 MHz or more is suitable for materials, solids and low impedance systems.
- Maximum current and compliance: battery and electrolyzer cells require larger currents, sometimes with an amplifier (booster).
- Low impedance accuracy: essential for batteries and supercapacitors; The 4-terminal measurement reduces the cable error.
- Number of channels: multi-channel instruments accelerate aging studies with multiple cells in parallel.
- Software: adjustment of equivalent circuits, Kramers-Kronig analysis and data export.
Good Practices
- Ensure the system is stable (constant open circuit potential) before measuring.
- Use small amplitudes to maintain the linear response.
- Use a Faraday cage for low current measurements.
- Keep cables short and braided to reduce inductance.
A LabSolutions offers potentiostats with EIS of lines Admiral Instruments (Squidstat), Sciospec, ALS and Corrtest, in addition to impedance tomography systems. See also the electrochemistry Catalog.
FAQ
What is the difference between potentiostatic and galvanostatic EIS?
In the potentiostatic, a potential disturbance is applied and the current is measured; In Galvanostatics, the opposite. Galvanostatic is preferred for batteries and low-current impedance systems.
Which frequency range should I use?
A 100 kHz to 10 mHz scan covers most electrochemical processes. Lower frequencies reveal diffusion but increase the measurement time.
What is the Kramers-Kronig test?
It is a mathematical check that indicates whether the impedance data is linear, causal and stable, that is, whether the measurement is reliable.